Vue de la page 65
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 66
Attribute Name Tag Type Notes
Modality (0008,0060) 1 MR/CT
Series Instance UID (0020,000E) 1 Locally Generated UID
Series Number (0020,0011) 2 Generated sequentially, always sent.
Laterality (0020,0060) 2C Always sent zero-length.
Series Date (0008,0021) 3 Generated for each series and always sent.
Series Time (0008,0031) 3 Generated for each series and always sent.
Performing Physicians Name (0008,1050) 3 Not Used. Discarded.
Protocol Name (0008,1030) 3 Sent if entered at user interface. Limited to
24 characters.
Series Description (0008,103E) 3 Value from user interface is sent.
Operators Name (0008,1070) 3 Value from user interface is sent. Limited
to 4 characters.
Referenced Study Component
Sequence
(0008,1111) 3 Not Used. Discarded
>Referenced SOP Class UID (0008,1150) 1C Value used for performed procedure step..
>Referenced SOP Instance UID (0008,1155) 1C Value used for performed procedure step..
Body Part Examined (0018, 0015) 3 Not Used. Will be discarded.
Patient Position (0018,5100) 2C Sent. As selected by operator when patient
is positioned:
Defined terms are:
HFP = Head-First Prone
HFS = Head-First Supine
HFDR = Head-First Decubitus Right
HFDL = Head-First Decubitus Left
FFDR = Feet-First Decubitus Right
FFDL = Feet-First Decubitus Left
FFP = Feet-First Prone
FFS = Feet-First Supine
Smallest Pixel Value in Series (0028,0108) 3 Not Used. Will be discarded.
Lardest Pixel Value in Series (0028,0109) 3 Not Used. Will be discarded.
A.5.5
General Equipment Module
Attribute Name Tag Type Notes
Manufacturer (0008,0070) 2 Always sent as GE MEDICAL
SYSTEMS
Institution Name (0008,0080) 3 Sent. Value is configurable. Limited to 32
characters.
Institution Address (0008,0081) 3 Not Used. Discarded.
Station Name (0008,1010) 3 Sent. Value is configurable. Limited to 8
characters.
Institution Department Name (0008,1040) 3 Not Used. Discarded.
Manufacturers Model Name (0008,1090) 3
Device Serial Number (0018,1000) 3 Sent if present in image.
Software Versions (0018,1020) 3 Always sent as 07 for this version - does
not distinguish individual software
releases.
Spatial Resolution (0018,1050) 3 Sent only for MR images.
Date of Last Calibration (0018,1200) 3
Time of Last Calibration (0018,1201) 3
Pixel Padding Value (0028,0120) 3 Sent. For CT images the value used may
not correspond to the actual value used in